题名 |
定量超音波骨密度儀於社區護理篩檢活動使用之可行性 |
并列篇名 |
Evaluation of Osteoporosis by Quantitative Ultrasound in Community Health Nursing for Screen |
DOI |
10.6974/TCNJ.200206.0056 |
作者 |
彭台珠(Tai-Chu Peng);李明憲(Ming-Shinn Lee);張芙美(Fwu-Mei Chang);黃森芳(Sen-Fang Huang);張惠君 |
关键词 |
骨密度 ; 骨質疏鬆症 ; 社區護理 ; 定量超音波 ; osteoporosis ; bone density ; community health nursing ; quantitative ultrasound QUS |
期刊名称 |
志為護理-慈濟護理雜誌 |
卷期/出版年月 |
1卷2期(2002 / 06 / 01) |
页次 |
56 - 63 |
内容语文 |
繁體中文 |
中文摘要 |
The purposes of this study were to understand and compare the precision and accuracy of quantitative ultrasound (QUS) with dual-energy x-ray absorptiometry (DEXA). First, the study was conducted from April 14, 2001 and ended on March 26, 2002, for internal machine quality assurance. Second, a total 60 healthy female workers participated in the random variation of two QUS tests. Third, 50 female members of a community fitness club took the QUS and DEXA of the lumbar spine. There were 30 results in the quality assurance test with a stable degree of stable instrument precision Variations of the two QUS tests for calcaneus were lower. The mean and stander deviation were -0.85±4.72, the 95% confidence interval was -2.07~0.37, pare T test were -1.40 p=0.17, Spearman's correlation=0.975 p<0.01. The correlation QUS and DEXA of lumbar spine were 0.521 p<0.01, which shows the validity of QUS is acceptable for community screening of lower bone density. |
英文摘要 |
The purposes of this study were to understand and compare the precision and accuracy of quantitative ultrasound (QUS) with dual-energy x-ray absorptiometry (DEXA). First, the study was conducted from April 14, 2001 and ended on March 26, 2002, for internal machine quality assurance. Second, a total 60 healthy female workers participated in the random variation of two QUS tests. Third, 50 female members of a community fitness club took the QUS and DEXA of the lumbar spine. There were 30 results in the quality assurance test with a stable degree of stable instrument precision Variations of the two QUS tests for calcaneus were lower. The mean and stander deviation were -0.85±4.72, the 95% confidence interval was -2.07~0.37, pare T test were -1.40 p=0.17, Spearman's correlation=0.975 p<0.01. The correlation QUS and DEXA of lumbar spine were 0.521 p<0.01, which shows the validity of QUS is acceptable for community screening of lower bone density. |
主题分类 |
醫藥衛生 >
預防保健與衛生學 醫藥衛生 > 社會醫學 |