题名

定量超音波骨密度儀於社區護理篩檢活動使用之可行性

并列篇名

Evaluation of Osteoporosis by Quantitative Ultrasound in Community Health Nursing for Screen

DOI

10.6974/TCNJ.200206.0056

作者

彭台珠(Tai-Chu Peng);李明憲(Ming-Shinn Lee);張芙美(Fwu-Mei Chang);黃森芳(Sen-Fang Huang);張惠君

关键词

骨密度 ; 骨質疏鬆症 ; 社區護理 ; 定量超音波 ; osteoporosis ; bone density ; community health nursing ; quantitative ultrasound QUS

期刊名称

志為護理-慈濟護理雜誌

卷期/出版年月

1卷2期(2002 / 06 / 01)

页次

56 - 63

内容语文

繁體中文

中文摘要

The purposes of this study were to understand and compare the precision and accuracy of quantitative ultrasound (QUS) with dual-energy x-ray absorptiometry (DEXA). First, the study was conducted from April 14, 2001 and ended on March 26, 2002, for internal machine quality assurance. Second, a total 60 healthy female workers participated in the random variation of two QUS tests. Third, 50 female members of a community fitness club took the QUS and DEXA of the lumbar spine. There were 30 results in the quality assurance test with a stable degree of stable instrument precision Variations of the two QUS tests for calcaneus were lower. The mean and stander deviation were -0.85±4.72, the 95% confidence interval was -2.07~0.37, pare T test were -1.40 p=0.17, Spearman's correlation=0.975 p<0.01. The correlation QUS and DEXA of lumbar spine were 0.521 p<0.01, which shows the validity of QUS is acceptable for community screening of lower bone density.

英文摘要

The purposes of this study were to understand and compare the precision and accuracy of quantitative ultrasound (QUS) with dual-energy x-ray absorptiometry (DEXA). First, the study was conducted from April 14, 2001 and ended on March 26, 2002, for internal machine quality assurance. Second, a total 60 healthy female workers participated in the random variation of two QUS tests. Third, 50 female members of a community fitness club took the QUS and DEXA of the lumbar spine. There were 30 results in the quality assurance test with a stable degree of stable instrument precision Variations of the two QUS tests for calcaneus were lower. The mean and stander deviation were -0.85±4.72, the 95% confidence interval was -2.07~0.37, pare T test were -1.40 p=0.17, Spearman's correlation=0.975 p<0.01. The correlation QUS and DEXA of lumbar spine were 0.521 p<0.01, which shows the validity of QUS is acceptable for community screening of lower bone density.

主题分类 醫藥衛生 > 預防保健與衛生學
醫藥衛生 > 社會醫學