题名

降低血液透析逆滲透水處理系統與透析液菌落數之專案

并列篇名

A Project for Reducing Reverse Osmosis Water Conditioning Systems and Bacteria Count in Hemodialysate

DOI

10.3966/172674042016031501002

作者

余秀芬(Hsiu-Fen Yu);張麗娟(Li-Chuan Chang)

关键词

病人安全 ; 透析液 ; 菌落數 ; patient safety ; dialysate ; bacteria count

期刊名称

臺灣腎臟護理學會雜誌

卷期/出版年月

15卷1期(2016 / 03 / 01)

页次

11 - 25

内容语文

繁體中文

中文摘要

透析醫療品質的首要重心是透析液,本單位於2010年4月至12月透析液菌落數偏高,確保病人安全故成立專案小組,旨在降低逆滲透(reverse osmosis, RO)水處理系統與透析液菌落數、提升透析護理品質。經分析發現問題有:透析機消毒時間不足、RO水處理系統採水時取樣口污染、透析機RO水入口處未執行無菌技術採水、人員對透析液品質知識不足、透析機老舊。改善方法為:(一)修訂透析室採水標準技術;(二)修訂水質異常處理作業規範、制訂設備管理作業規範、制訂RO設備消毒作業規範;(三)更換老舊血液透析機;(四)建立稽核機制;(五)舉辦在職教育等方法,結果菌落數異常件數3.92件/月降至0件/月,顯示病人安全得到保障也提昇了透析品質。

英文摘要

Dialysate is regarded as the key in the healthcare quality of dialysis. The elevated dialysate bacteria count of this unit between April and December of 2010 led to the establishment of an ad hoc group assigned to reducing the bacteria count in the reverse osmosis (RO) water conditioning systems and dialysate, as well as enhancing the quality of nursing care during dialysis. Some of the problems discovered through the analysis included inadequate sterilization of dialysis machines, sampling port pollution of RO water conditioning systems, failure to use aseptic water intake techniques for the RO water inlets of dialysis machines, insufficient knowledge about dialysate quality, and worn dialysis machines. Suggested improvements included: (1) revising the standard technique for water intake in dialysis units; (2) revising the standard operating procedures for handling abnormal water quality, equipment management, and RO apparatus disinfection; (3) replacement of worn hemodialysis machines; (4) establishing auditing mechanisms, and (5) hosting of inservice education. The number of cases of abnormal bacteria count subsequently dropped from 3.92 to 0 cases per month. Patient safety was therefore assured while dialysis quality was improved as well.

主题分类 醫藥衛生 > 內科
醫藥衛生 > 社會醫學
参考文献
  1. Borges, C. R. M.,Lascowski, K. M. S.,Filho, N. R.,Pelayo, J. S.(2007).Microbiological quality of water and dialysate in a haemodialysis unit in Ponta Grossa-PR Brazil.Journal of Applied Microbiology,103(5),1791-1797.
  2. Chamney, M. J.,James, R. C.(2008).Dialysis water quality for renal nurses.Renal Society of Australasia Journal,4(1),13-20.
  3. Hoenich, N. A.,Levin, R.,Ronco, C.(2010).Warter for haemodialysis and related therapies: Recent standards and emerging issues.Blood Purify,29(2),81-85.
  4. Montanari, L. B.,Sartori, F. G.,Cardoso, M. J.,Varo, S. D.,Pires, R. H.,Leite, C. Q.,Martins, C. H.(2009).Microbiological contamination of a hemodialysis center water distribution system.Revista do Instituto de Medicina Tropical de Sao Paulo,51(1),37-43.
  5. Varo, S. D.,Martins, C. H.,Cardoso, M. J.,Sartori, F. G.,Montanari, L. B.,Pires-Goncalves, R. H.(2007).Isolation of filamentous fungi from water used in a hemodialysis unit.Revista da Sociedade Brasileira Medicina Tropical,40(3),326-331.
  6. 胡連順,黃正誼(2009)。如何維持透析液的品質。腎臟與透析,21(1),63-67。
  7. 徐宗福(2007)。根本原因分析在醫療照護的應用。護理雜誌,54(6),77-83。
  8. 張炎林,劉紹毅(2008)。醫療單位監督與維持血液透析治療用水規範。血液淨化雜誌,13(1),47-64。
  9. 張炎林,劉紹毅(2008)。血液透析治療用水處理設備標準。血液淨化雜誌,13(1),25-45。
  10. 游志清,徐永堂,鄭振廷(2006)。從AAMI 2005 談血液透析用液規範。血液淨化雜誌,11(1),45-61。
  11. 楊漢湶(2009)。2008-2009 病人安全年度目標介紹。醫療品質雜誌,2(5),7-11。
  12. 劉紹毅,陳再晉,林梅圃,張琦玲,劉才睿(2007)。內毒素 LAL 檢測在透析機維護的應用。血液淨化雜誌,12(1-2),71-78。
  13. 鄭輝賢(2008)。血液透析室之 RO 管路消毒經驗分享。腎臟與透析,20(2),121-123。
被引用次数
  1. 崔若畇,姜美珠,吳春桂(2019)。降低血液透析室逆滲透水菌落數異常率之專案。臺灣腎臟護理學會雜誌,18(1),31-45。