题名

紙片電噴灑質譜法的特性與其作為快速篩選尿液中氯安非他命之研究

并列篇名

Rapid Screening and Determination of Chloroamphetamines in Urine by a Paper Spray-Mass Spectrometry

DOI

10.6134/tjfm.2013.0502.03

作者

陳昕楷(Hsin-Kai Chen);李珣(Hsun Lee);張智勝(Chih-Sheng Jhang);陳姵君(Pei-Chun Chen);柳如宗(Ju-Tsung Liu);林震煌(Cheng-Huang Lin)

关键词

紙片電噴灑質譜法 ; 快速篩選 ; 尿液 ; 氯安非他命 ; Paper spray ; Rapid screening ; Urine ; Amphetamines

期刊名称

台灣法醫學誌

卷期/出版年月

5卷2期(2013 / 12 / 01)

页次

16 - 23

内容语文

繁體中文

中文摘要

本文介紹紙片筆尖協助-紙片電噴灑質譜的特性與其作為快速篩選尿液中氯安非他命之研究。實驗結果發現,紙片尖端的角度為15度時,高壓電為三千伏特時,紙片尖端與質譜入口的最佳距離為7.7 mm。此條件下可得最佳偵測極限。當o-,m-,p-氯安非他命及o-,m-,p-氟安非他命為測試樣品時,偵測極限可達0.1 μg/mL。本研究亦使用傳統的快速篩選質譜法,包括了大氣基質輔助雷射脫附游離質譜法與電噴灑輔助雷射脫附游離法。實驗結果發現紙片電噴灑法不僅在靈敏度高出其他兩種方法,而且使用的設備簡單,完全不需要較昂貴的雷射等設備。

英文摘要

A method for the rapid screening and determination of amphetamine-type designer drugs in urine by a novel nibassisted paper spray-mass spectrometry (NAPS-MS) procedure is described. When a sharp (15°-tip) chromatography paper was used, the optimized distance from the paper tip to the mass inlet was 7.7 mm. Under optimized conditions, the limit of detections for amphetamine derivatives (model samples: o-, m-, p-chloro- and o-, m-, p-fluoro-amphetamines, respectively) were determined to 0.1 μg/mL by the NAPS-MS method. This method is easier and has a higher sensitivity than similar methodologies, including atmospheric pressure/matrix-assisted laser desorption ionization mass spectrometry and electrospray-assisted laser desorption ionization/mass spectrometry.

主题分类 醫藥衛生 > 基礎醫學
社會科學 > 法律學
参考文献
  1. Bahr, U,Deppem, A,Karas, M,Hillenkamp, F,Giessmann, U(1992).mass spectrometry of synthetic polymers by UV-matrix-assisted laser desorption/ionization.Anal Chem,64,2866-9.
  2. Cooks, RG,Manicke, NE,Dill, AL(2011).New ionization methods and miniature mass spectrometers for biomedicine: DESI imaging for cancer diagnostics and paper spray ionization for therapeutic drug monitoring.Faraday Discuss,149,247-67.
  3. Espy, RD,Manicke, NE,Ouyang, Z,Cooks, RG(2012).Rapid analysis of whole blood by paper spray mass spectrometry for point-of-care therapeutic drug monitoring.Analyst,137,2344-9.
  4. Espy, RD,Muliadi, AR,Ouyang, Z,Cooks, RG(2012).Spray mechanism in paper spray ionization.Int J Mass Spectrom,325-27,167-71.
  5. Grasso, G,Fragai, M,Rizzarelli, E,Spoto, G,Yeo, KJ(2006).In situ AP/MALDI-MS characterization of anchored matrix metalloproteinases.J Mass Spectrom,41,1561-9.
  6. Grasso, G,Mineo, P,Rizzarelli, E,Spoto, G(2009).MALDI, AP/MALDI and ESI techniques for the MS detection of amyloid beta-peptides.Int J Mass Spectrom,282,50-5.
  7. Huang, MZ,Hsu, HJ,Wu, CI(2007).Characterization of the chemical components on the surface of different solids with electrospray-assisted laser desorption ionization mass spectrometry.Rapid Commun Mass Spectrom,21,1767-75.
  8. Huang, MZ,Jhang, SS,Cheng, CN,Cheng, SC,Shiea, J(2010).Effects of matrix, electrospray solution, and laser light on the desorption and ionization mechanisms in electrospray-assisted laser desorption ionization mass spectrometry.Analyst,135,759-66.
  9. Ifa, DR,Wu, C,Ouyang, Z,Cooks, RG(2010).Desorption electrospray ionization and other ambient ionization methods: current progress and preview.Analyst,135,669-81.
  10. Liu, JJ,Wang, H,Manicke, NE,Lin, JM,Cooks, RG,Ouyang, Z(2010).Development, characterization, and application of paper spray ionization.Anal Chem,82,2463-71.
  11. Manicke, NE,Yang, QA,Wang, H,Oradu, S,Ouyang, Z,Cooks, RG(2011).Assessment of paper spray ionization for quantitation of pharmaceuticals in blood spots.Int J Mass Spectrom,300,123-9.
  12. Morelato, M,Beavis, A,Kirkbride, P,Roux, C(2013).Forensic applications of desorption electrospray ionisation mass spectrometry (DESI-MS).Forensic SciInt,226,10-21.
  13. Peng, IX,Little, MW,Loo, RRO,Margalith, E,Loo, JA(2010).Ambient electrospray-assisted laser desorption Ionization (ELDI) for mass spectrometry characterization of proteins and glycoproteins.Abstr Pap Am Chem Soc,240.
  14. Peng, IX,Loo, RRO,Margalith, E,Little, MW,Loo, JA(2010).Electrospray-assisted laser desorption ionization mass spectrometry (ELDI-MS) with an infrared laser for characterizing peptides and proteins.Analyst,135,767-72.
  15. Perdian, DC,Schieffer, GM,Houk, RS(2010).Atmospheric pressure laser desorption/ionization of plant metabolites and plant tissue using colloidal graphite.Rapid Commun Mass Spectrom,24,397-402.
  16. Shiea, J,Huang, MZ,Hsu, HJ(2005).Electrospray-assisted laser desorption/ionization mass spectrometry for direct ambient analysis of solids.Rapid Commun Mass Spectrom,19,3701-4.
  17. Tanaka, K,Waki, H,Ido, Y(1988).Protein and polymer analyses up to m/z 100,000 by laser ionization time-of-flight mass spectrometry.Rapid Commun Mass Spectrom,2,151-3.
  18. Vickerman, JC(2011).Molecular imaging and depth profiling by mass spectrometry-SIMS, MALDI or DESI?.Analyst,136,2199-217.
  19. Wang, H,Liu, JJ,Cooks, RG,Ouyang, Z(2010).Paper spray for direct analysis of complex mixtures using mass spectrometry.Angew Chem-Int Edit,49,877-80.