题名 |
Stability-Indicating High-Performance Liquid Chromatography Assay for the Determination of Sulthiame in Pharmaceutical Dosage Forms |
DOI |
10.4137/ACI.S38656 |
作者 |
Ammar Haidar;Sofiane Kabiche;Elyes Majoul;Issa-Bella Balde;Jean-Eudes Fontan;Salvatore Cisternino;Joël Schlatter |
关键词 |
sulthiame ; liquid chromatography ; stability indicating ; forced degradation |
期刊名称 |
Analytical Chemistry Insights |
卷期/出版年月 |
2016(2016 / 12 / 01) |
页次 |
59 - 64 |
内容语文 |
英文 |
中文摘要 |
A stability-indicating assay by reversed-phase high performance liquid chromatography method was developed and validated for the determination of sulthiame (STM). The chromatographic separation was achieved on a reversed-phase NovaPack C18 column and an isocratic mobile phase consisting of deionized water:methanol (70:30, v/v). The flow rate was 1.0 mL/min (ultraviolet detection at 210 nm). The STM was separated within 2.83 min. The linearity of the method was demonstrated in the range of 20.0-200.0 μg/mL and a coefficient of determination of r^2 = 0.9999. The limits of detection and quantification were 4.2 and 9.5 μg/mL, respectively. The intraday and interday precisions were less than 1%. Accuracy of the method ranged from 98.3% to 101.7%, with a relative standard deviation of <1%. STM was degraded by accelerated breakdown in alkaline, acidic, or oxidative stress conditions. This method allows accurate and reliable determination of STM for drug stability assay in pharmaceutical studies. |
主题分类 |
基礎與應用科學 >
化學 |